Experiments quiz This is a short quiz about the experimental set-up of experiments. It will take only a couple of minutes to complete. Succes! Question 1: Double referencing Why should you use double referencing? to correct for ligand versus reference differences to correct for blank (buffer only) injections to correct for bulk differences in the analyte to correct for drift during the analyte injection Question 2: Kinetic inject Why should you use the 'Kinetic / high performance' inject method? to separate the sample from the flow buffer to have slow injections to have lower bulk shift jumps to have better dissociation curves Question 3: Slow dissociation With a slow dissociation, the best set up is to use multi cycle kinetics single cycle titration kinetics short & long kinetics steady state kinetics Question 4: Randomization Randomization of the samples is done because it minimizes systematic errors it can reveal systematic errors it proves the system is reproducible it minimizes carry over between samples Question 5: Kinetic estimation To get a better estimation of the kinetics you repeat the method on the same sensor chip use a wide analyte concentration range add more analyte injections in the method repeat the method with different ligand and analyte batches Question 6: Mass transport How do you detect mass transport limitation? by immobilizing less ligand by varying the flow rate during analyte injection by injection several analyte concentrations by varying the analyte injection time Question 7: Non specific What can you try to minimize non-specific binding? use a low pH running buffer use a higher flow rate lower the analyte concentration adding more salt or detergent to the running buffer Question 8: Baseline drift What can you do when there is baseline drift? use a higher flow rate immobilize less ligand flow for a longer period of time with flow buffer ad more salt or detergent to the running buffer Question 9: Concentration range What is the best analyte concentration range to inject? 0 – 1 µM 0 – 100% of Rmax 0.1 – 10 times the expected KD at least one concentration should saturate the ligand Question 10: Replicate Replicate analyte injections are done to prove the system is stable and reproducible to have more curves to analyse to reveal dilution errors to minimize systematic errors Question 11: Replicates Why should replicates be in separate capped vials? to minimize sample dilution to minimize sample evaporation to minimize carry over between samples to minimize sample loss to the vial wall Question 12: Calibration Some SPR instruments use a calibration routine. Why? to initialyse the fluidics to get the best detector sensitivity to make sure the curves are separated well to make sure the system is equilibrated Question 13: Statements Which of the statements is not true? A slow dissociation needs a longer dissociation time than a fast dissociation to resolve the kinetics To determine the Rmax, the ligand has to be fully saturated SPR can be used for both high and low affinity interactions Both the association and dissociation curve follow an exponential curve Question 14: Buffers Why should the buffers be filtered and degassed before use? To prevent air spikes and drift To minimize the bulk effect To have a better buffer equilibration To have a better 1:1 interaction Question 15: Kinetics To determine the ka and kd of an interaction you need at least One analyte injection with 1x KD concentration + dissociation One analyte injection that is saturating the ligand + dissociation Two analyte injections to steady state + dissociation Two analyte injections with curvature + dissociation Details Last Updated: 08 August 2022